HPLC Procedure for DNA
Purification
TURNING ON HPLC
- Press "sleep" buttons to
wake up machines.
- Transfer PUMP B from 100% ACN jar to
buffer B jar.
- Turn on lamp
a.
In "Analysis" screen, set "Lamp" to "1".
b.
Press "zero" button on SPD-10A.
- On computer:
a.
Open "CLASS VP Chromatography".
b.
Open "Corn HPLC".
Normally
the program is already opened in the computer.
WASHING THE
COLUMN
- In "Analysis File" screen
(1 in main
menu), set "conc" to "100%", then press
"PMP.ON" button (f1).
This flows 100% buffer B to the column, which takes 15-20 min to
equilibrate.
- After pressure is stable, set
"conc" to "10%".
This flows 10% buffer B and 90% buffer A. When the pressure is
stable, the HPLC is ready for sample injection.
INJECTING
THIOL-DNA
- In "sequence" screen (press
0 in main menu)
a.
Set 'sample #' = 1 - 1
b.
Set volume slightly less than sample volume, for example, 1 uL less
c.
Set 'file' = F0
- Put sample vial in position
"1" on the sample rack.
- Injection:
- Press "single" on the
computer, enter directory and file name
- On system controller (SCL-10A),
press "run" button
- Wait for ~20 min for the DNA peak to
show up on the computer screen.
- After collecting the DNA fraction,
click on the "stop" button on the computer, and then press the
"run" button on the system controller (SCL-10A) to stop HPLC in
the middle of the run.
- If you have a second sample to
purify, you need to wash the column by changing ̉conÓ to 100% and allowing
the instrument to equilibrate.
Meanwhile, prepare your second sample for HPLC. After a stable pressure has been
reached, change the ̉concÓ to 10%, allow to stabilize and put your second
vial in position ̉1Ó. Repeat
steps from # 1 in this section.
SHUTTING DOWN
1. Turn off
lamp: set "Lamp" to 0 in "analysis" screen.
2. Turn off
pump:
a.
In "analysis" screen, press "PMP.OFF" button (f1).
b.
Wait until pump pressure drop to zero, then transfer pump B to 100% ACN.
c.
Turn pump on again, until pressure equilibrates (600-700 psi for ACN).
d.
Then turn pump off. Make sure pressure drops to zero.
3. Press the
"sleep" button to put machines to sleep (DO NOT turn off the power).